500K Hours based on Telecordia SR-332 (at 115Vac, Max. load and 25°C Ambient)
Expected Cap Life Time
3 years (115 Vac, 100% load @ 25°C Ambient)
CE
In conformance with EMC Directive 2014/30/EU and Low Voltage Directive 2014/35/EU EN 60601-1: 2006 + A1: 2013 + A12: 2014+ A2: 2021 & EN 60601-1-2: 2015 + A1: 2021
UKCA
In conformance with Electrical Equipment (Safety) Regulations 2016 and Electromagnetic Compatibility Regulations 2016, Medical Devices Regulations 2002 (UK MDR 2002)
Harmonic Current Emissions
IEC 61000-3-2 Meet Class A
Voltage Fluctuation and Flicker
IEC 61000-3-3
Line Regulation
± 0.5%
Load Regulation
± 5%
Start-up Time
350 ms typ. @ 115 Vac & 230 Vac
Hold-up Time
28 ms typ. @ 115 Vac
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Overvoltage
150% max, Latch Mode
Overload / Overcurrent
180% max of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery)
Over Temperature
Latch Mode
Short Circuit
Hiccup Mode, (Non-Latching, Auto-Recovery)
Protection Against Shock
Class I
Leakage Current
0.1 mA @ 264 Vac NC , 0.1 mA @ 264 Vac SFC
Operating Altitude
Up to 5,000 meters (up to 16,400 feet or 106-54kPa)