Related products
The process calibrator is also a mains-free voltage calibrator used to simulate control or unit signals in the MSR technique, with which you can check and calibrate almost all parameters of units and measuring devices. The calibrator is therefore used to provide an electrical output signal (mA, mV, V and frequency in Hz as well as temperature in °C or °F). This voltage calibrator also has automatic leveling or graduation functions, a percentage selection of the output signal jumps (from 0 to 100%) by means of the up/down keys and a direct indication of the output value (p.e.13,8mA). It is intuitively and directly operated via the keyboard. This process calibrator is characterized by its high precision and universal range of use, which is why it should not be missing in any electrical workshop or MSR department, as well as in industry or automation technology.
- 4 ... 20 mA (1 kΩ, 24 V power)
- 0 ... 100.00 mV / 0 ... 1.0000 V / 0 ... 12,000V
- K, J, E, T temperature sensor (°C and °F)
- Frequency Range 1 ... 62500 Hz
- Basic accuracy 0.025%
- Easy operation with the keyboard
- Graduated automatic functions
- To choose between 0 ... 20 mA/ 0 ... 24 mA
- Input 0 ... 100% (mA, mV, V)
- "Open" (mA) or "short circuit" (mV, V) warning beep
Calibrator for simulation of electrical signals as well as frequency and temperature | |
Technical Specifications | |
Current mA DC (maximum load at 1 kΩ, power at 24 V) | |
Rank | 4 ... 20 mA/ 0 ... 20 mA/ 0 ... 24 mA |
Resolution | 1 μA |
Precision | ± 0.025% ± 3 μA |
Notice | Beep in case of short circuit or current output > 1 mA |
Current mV, V DC (power supply 1 mA) | |
Ranges | 0 ... 100.00 mV 0 ... 10,000 V 0 ... 1.0000 V |
Resolution | 10 μV 1 mV 1.0000 V |
Precision | ± 0.05% ± 30 μV ± 0.05% ± 3 mV ± 0.05% ± 300 μV |
Notice | Beep in case of short circuit or current output > 10 mV. |
Temperature sensor type K, J, E, T ; (1 kΩ min.) | |
Ranges | K:-200 ... 0 °C and 0 ... + 1370 °C ; J: ; - 100 ... 0 °C and 0 ... + 760 °C E:-100 ... 0 °C and 0 ... + 700 °C T: ; - 200 ... 0 °C and T: 0 ... + 400 °C |
Resolution | 1°C (for everyone) |
Precision | K: ± 1.1 °C; and ± 0.8 °C J: ; ± 0.9 °C and ± 0.7 °C E: ± 0.9 °C and ± 0.7 °C T: ; ± 1.0 °C and ± 0.8 °C |
Frequency | |
Rank | 1 ... 125 Hz and 126 ... 62.5 kHz |
Resolution | 1... 125 Hz / 1 Hz 126...62.5 kHz / 604 steps |
Precision | ±0.04 Hz |
Selection | 604 frequencies |
General characteristics | |
Screen | 5-position LCD |
Operating Temperature Range | 0 to +50 °C |
Operating Humidity Range | |
Storage Temperature | -20 to +60 °C |
Storage Humidity | |
Feeding | 1 x 9V battery or ; 6 x 1.5V |
Consumption | 60 mA ... 180 mA (depending on output) |
Battery indicator | 5.5 V for 150 mA |
Dimensions | 88 x 168 x 26 mm |
Weight | 330 g |
What's in the Shipment 1 x Process Calibrator PCE-123 1 x Briefcase 1 x Type K Adapter 1 x Battery Holder 6 x Batteries 1x Test cables with alligator clips 1 x Instructions for Use |
No customer reviews for the moment.